Product.
Detection
HiSense™ Mycoplasma PCR Detection Kit
This kit is designed to detect mycoplasma contamination in production cell lines and biopharmaceuticals through conventional PCR with high sensitivity.
HD-100, HDEP-100
Ordering Information
Cat. No | Unit size | Price |
HD-100 | 100 Tests | Request a quote→ |
HDEP-100 | 100 Tests |
Product Overview
Features
- Mycoplasma test in production cell lines and biopharmaceuticals(Compliance with Korea MFDS, EP and JP)
- Detection of 209 types of mycoplasma with high sensitivity
- Template DNA: Boiling of culture supernatant or genomic DNA extraction from cultured cells
- High Sensitivity: 1~10 copies/reaction
- Specificity : Do not detect other bacterial species with close phylogenetic relation to mycoplsma
- UDG system : Prevention of carry-over contamination
Results
Gel Electrophoresis Result
Tested with Quantitative M. arginini Genomic DNA (Cat. No. QAGD, CellSafe)
Lane 1: 104 copies
Lane 2: 103 copies
Lane 3: 102 copies
Lane 4: 101 copies
Lane 5: 1 copy
Lane 6: Positive Control
Lane 7,8: Negative Control
Lane 9: 100bp ladder
- When mycoplasma contamination exists, a band with around 250-300 bp appears.